Pressure drops on ACQUITY UPLC with i2V valves - WKB221275
Article number: 221275
SYMPTOMS
- Pressure drops
- Pump leak test passes
- Replacement of solvent filter does not solve problem
- Replacement of primary and accumulator check valves does not solve problem
ENVIRONMENT
- ACQUITY UPLC
- ACQUITY UPLC BSM
- ACQUITY UPLC I-Class
- ACQUITY UPLC I-Class PLUS
CAUSE
Clogged filter frit in the Filter Cartridge Assembly
FIX or WORKAROUND
- Remove the filter cartridge assembly from i2V valve.

2. Sonicate the filter frit in appropriate solvent (for example, water, methanol, isopropanol).

3. If it is impossible to clean the filter frit, replace it.
ADDITIONAL INFORMATION
Part number for filter frit (same number for ACQUITY UPLC I-Class and ACQUITY UPLC): 700002913