Skip to main content
Waters

ACQUITY UPLC fails needle seal leak test - second solution - WKB220730

Article number: 220730

SYMPTOMS

  • ACQUITY UPLC fails needle seal leak test

ENVIRONMENT

  • ACQUITY UPLC (Classic) Sample Manager

CAUSE

Stainless steel needle leaks.

FIX or WORKAROUND

Replace the needle.

ADDITIONAL INFORMATION