Empower is not communicating with a Beckman PA 800 connected via e-SAT/IN - WKB21894
Article number: 21894
SYMPTOMS
- No communication with system
- e-SAT/IN reads "OK? - No" in the node properties
ENVIRONMENT
- Empower 3
- Beckman PA 800
- e-SAT/IN module
CAUSE
The Beckman PA 800 was not set to a static IP address.
FIX or WORKAROUND
Configure the Beckman unit to static IP.
ADDITIONAL INFORMATION