Skip to main content
Waters

FTN error: Missing/wrong plate or vial - WKB209335

Article number: 209335

SYMPTOMS

  • FTN error: Missing/wrong plate or vial

ENVIRONMENT

  • ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)

CAUSE

The sample plate is installed incorrectly. 

FIX or WORKAROUND

Ensure that the A1 position of the sample plate is in the upper left-hand corner.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.