Is it possible to change the gain settings using timed events on an ACQUITY UPLC FLR Detector to avoid detector saturation? - WKB209056
Article number: 209056
ENVIRONMENT
- ACQUITY UPLC FLR Detector
- Empower
ANSWER
Yes, this is possible and should be used as a standard practice during method development. Timed events are described in the ACQUITY UPLC FLR Detector
user manual. https://www.waters.com/webassets/cms/support/docs/71500142403rb.pdf
ADDITIONAL INFORMATION
To change PMT Gain on the FLR during the acquisition, do it in the instrument method's Timed Events table.
The channel field is unavailable because PMT Gain is per detector, not per channel.