"Z axis move h/w fault" and "Zp axis move h/w fault" errors upon injection on FTN - WKB208332

Article number: 208332

SYMPTOMS

  • Upon injection, one or both of the following errors appear: "Z axis move h/w fault", "Zp axis move h/w fault" 
  • Errors are seen in the error log in the ACQUITY console
  • Plate configuration is correct

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC I-Class 

CAUSE

Vial position typographical error or incorrect vial position format

FIX or WORKAROUND

Enter the correct vial position or format.

ADDITIONAL INFORMATION

 