Do I need to prime all Alliance solvent lines, even if I am not using some of them? - WKB206913

Article number: 206913

ENVIRONMENT

  • Alliance Separations Module
  • 2695
  • 2795

ANSWER

Yes, at least occasionally.

The degasser solvent chambers are connected to a single vacuum engine, which can efficiently degas solvent only if all of the chambers are "wet".

ADDITIONAL INFORMATION

See discussion in the Waters Discussion Group.