Compression check fails on Alliance 2695 Separations Module - WKB204692

Article number: 204692

SYMPTOMS

  • Compression check fails
  • New seal pack
  • New syringe
  • Injector purged several times
  • Solvent leaks on valve housing
  • Restriction missing

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 717plus Autosampler

CAUSE

The V1 valve (high-pressure valve) must be replaced.

FIX or WORKAROUND

Replace the V1 valve and/or the V2 valve (WAT045438 - High Pressure Motorized Valve Assembly).