Alliance 2695 actuator fails static leak test - WKB20218

Article number: 20218

SYMPTOMS

  • Actuator fails static leak test
    • "Accumulator unable to compress" error occurs
    • "Accumulator leak rate too high..." error occurs
  • Solvent is present around the face seals on the back of the accumulator J-tube head assembly

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

Worn face seals

FIX or WORKAROUND

1. Remove the accumulator head using the Head Removal/Replacement Diagnostic.

2. Replace the face seals, Face Seals Replacement Kit 4/Pkg [WAT270939].

3. Repeat the static leak test.

ADDITIONAL INFORMATION

id20218, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB