Article number: 202034

SYMPTOMS

  • "Sample pressure low (##): possibly no wash solvent" error appears intermittently.
  • Sample and wash syringe leak tests pass.
  • Needle seal leak test passes.
  • Back pressure regulator passes.
  • Needle seal and needle loop and volume characterizations pass.
  • No visible leaks at wash manifold.
  • When inspecting needle wash port, needle seal o-ring looks fine (p/n - 700002572)

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC I-Class

CAUSE

Worn threads on needle wash port or over use of teflon tape

FIX or WORKAROUND

  1. Remove excess teflon tape on threads of needle wash port and apply a fresh wrap or replace needle wash port (part number - 700002886)

ADDITIONAL INFORMATION