What is the correct volume of the 2998 Analytical Flow Cell? - WKB201039
Article number: 201039
ENVIRONMENT
- 2998
ANSWER
The volume of the analytical flow cell is listed as 9.3 µL in the 2998 Operator's Guide (released version 71500121902), but the specification sheet (720004519EN) gives the correct volume for the analytical flow cell—8.4 µL.
ADDITIONAL INFORMATION
JIRA CXKM-10027 requests a correction to the Operator's Guide.