Alliance HPLC System fails injector linearity and accuracy test - WKB200286
Article number: 200286
SYMPTOMS
- Successful linearity test with R^2 > or = 0.9990
- Accuracy test failure indicates a problem with the y-intercept or slope of the line
- Replacing the injector assembly does not help
ENVIRONMENT
- Alliance HPLC System
- 2695 Separations Module
- 2795 Separations Module
CAUSE
Faulty syringe assembly
FIX or WORKAROUND
Replace the 250-µL syringe assembly (WAT073109).
ADDITIONAL INFORMATION
The faulty syringe demonstrated irregular resistance over the range of movement—possibly non-uniform surface in the bore of the glass barrel.
Wetting the PTFE plunger tip before inserting it into the glass barrel can mitigate wear to the tip.
id200286,