Skip to main content
Waters

Alliance HPLC System fails injector linearity and accuracy test - WKB200286

Article number: 200286

SYMPTOMS

  • Successful linearity test with R^2 > or = 0.9990
  • Accuracy test failure indicates a problem with the y-intercept or slope of the line
  • Replacing the injector assembly does not help

ENVIRONMENT

  • Alliance HPLC System
  • 2695 Separations Module
  • 2795 Separations Module

CAUSE

Faulty syringe assembly

FIX or WORKAROUND

Replace the 250-µL syringe assembly (WAT073109).

ADDITIONAL INFORMATION

The faulty syringe demonstrated irregular resistance over the range of movement—possibly non-uniform surface in the bore of the glass barrel.

Wetting the PTFE plunger tip before inserting it into the glass barrel can mitigate wear to the tip.

id200286,

Not able to find a solution? Click here to request help.