No data being acquired with ACQUITY PDA on MassLynx - WKB199174
Article number: 199174
SYMPTOMS
- No data being acquired
- Flat zero line seen in chromatogram
ENVIRONMENT
- ACQUITY PDA
- MassLynx
CAUSE
Partial loss of communication to PDA
FIX or WORKAROUND
- Power-off the PDA and the host PC (run stopacquityprocesses.exe before shutting down the PC).
- Restart the PC and PDA.