Carryover seen with Alliance 2695 - WKB19856

Article number: 19856

SYMPTOMS

  • Carryover observed

ENVIRONMENT

  • Alliance HPLC system
  • 2695 Separations Module

CAUSE

Kinked tubing

FIX or WORKAROUND

  1. Inspect the green needle-wash tubing and the yellow needle-wash waste lines for kinks.
  2. Using a tubing cutter, cut the tubing to remove kinks.

ADDITIONAL INFORMATION

If still failing, replace the needle wash pump (WAT078979)