ACQUITY Arc system fails dynamic leak test on paths 1 and 2 - WKB197908

Article number: 197908

SYMPTOMS

  • System fails leak test on paths 1 and 2 with excessive leak rate

ENVIRONMENT

  • ACQUITY Arc

CAUSE

Leak at the stainless steel tube between solvent and sample managers

FIX or WORKAROUND

Replace the tube (700010703 - TUBE. SST, .062 X 0.10 X 19L).

ADDITIONAL INFORMATION

Arc Bio version: 700012305 - TUBE, MP35N, .062 X 0.10 X 19L