How to set IEEE on a 2996 - WKB196723
OBJECTIVE or GOAL
Set the IEEE-488 address on a 2996 Detector.
ENVIRONMENT
- 2996 PDA Detector
PROCEDURE
To set the IEEE-488 address for the 2996 Detector: Use a small screwdriver (or similar device) to set the IEEE-488 switches on the detector rear panel to the IEEE-488 address of the 2996 Detector. The address must be a number from 2 to 29 and must be different than that of any other component connected to your acquisition server.
IEEE Address Settings
Set IEEE switches to “1” for ON and to “0” for OFF.
|
5 |
4 |
3 |
2 |
1 |
|
1 |
2 |
4 |
8 |
16 |
|
|||||
|
1 |
‑‑‑ |
‑‑‑ |
‑‑‑ |
‑‑‑ |
‑‑‑ |
|
2 |
OFF |
ON |
OFF |
OFF |
OFF |
|
3 |
ON |
ON |
OFF |
OFF |
OFF |
|
4 |
OFF |
OFF |
ON |
OFF |
OFF |
|
5 |
ON |
OFF |
ON |
OFF |
OFF |
|
6 |
OFF |
ON |
ON |
OFF |
OFF |
|
7 |
ON |
ON |
ON |
OFF |
OFF |
|
8 |
OFF |
OFF |
OFF |
ON |
OFF |
|
9 |
ON |
OFF |
OFF |
ON |
OFF |
|
10 |
OFF |
ON |
OFF |
ON |
OFF |
|
11 |
ON |
ON |
OFF |
ON |
OFF |
|
12 |
OFF |
OFF |
ON |
ON |
OFF |
|
13 |
ON |
OFF |
ON |
ON |
OFF |
|
14 |
OFF |
ON |
ON |
ON |
OFF |
|
15 |
ON |
ON |
ON |
ON |
OFF |
|
16 |
OFF |
OFF |
OFF |
OFF |
ON |
|
17 |
ON |
OFF |
OFF |
OFF |
ON |
|
18 |
OFF |
ON |
OFF |
OFF |
ON |
|
19 |
ON |
ON |
OFF |
OFF |
ON |
|
20 |
OFF |
OFF |
ON |
OFF |
ON |
|
21 |
ON |
OFF |
ON |
OFF |
ON |
|
22 |
OFF |
ON |
ON |
OFF |
ON |
|
23 |
ON |
ON |
ON |
OFF |
ON |
|
24 |
OFF |
OFF |
OFF |
ON |
ON |
|
25 |
ON |
OFF |
OFF |
ON |
ON |
|
26 |
OFF |
ON |
OFF |
ON |
ON |
|
27 |
ON |
ON |
OFF |
ON |
ON |
|
28 |
OFF |
OFF |
ON |
ON |
ON |
|
29 |
ON |
OFF |
ON |
ON |
ON |
|
30 |
‑‑‑ |
‑‑‑ |
‑‑‑ |
‑‑‑ |
‑‑‑ |
ADDITIONAL INFORMATION