Skip to main content
Waters

ACQUITY SM x-axis and z-axis errors - WKB193536

Article number: 193536

SYMPTOMS

  • X-axis home h/w fault
  • Z-axis move h/w fault

ENVIRONMENT

  • ACQUITY UPLC Sample Manager

CAUSE

Y carriage collided with vial holder.

FIX or WORKAROUND

  1. Remove the needle and inspect for damage. Replace if damaged or refit if OK (default PEEK 30-µL sample needle = 700002644). 
  2. Calibrate XY and Zp axes.
  3. Calibrate Zp offset.
  4. Characterize needle seal.
  5. It is good practice to re-characterize the needle and loop volumes.

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.