ACQUITY SM x-axis and z-axis errors - WKB193536
Article number: 193536
SYMPTOMS
- X-axis home h/w fault
- Z-axis move h/w fault
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Y carriage collided with vial holder.
FIX or WORKAROUND
- Remove the needle and inspect for damage. Replace if damaged or refit if OK (default PEEK 30-µL sample needle = 700002644).
- Calibrate XY and Zp axes.
- Calibrate Zp offset.
- Characterize needle seal.
- It is good practice to re-characterize the needle and loop volumes.
ADDITIONAL INFORMATION
id193536, UPBSMIVD, UPIBSMIVD, UPLCCHIVD, UPLCCOLHTR, UPNBINARY, UPSMIVD, UPSMPMGR, UPSMPMGR