ACQUITY UPLC BSM dynamic leak test failure - WKB19234
Article number: 19234
SYMPTOMS
- Dynamic leak test fails to pressurize or compress
- When compression occurs, the accumulator actuator suffers very high leaks and the test fails to complete
- Pump heads rebuilt with pump head/seal wash housings, check valves, plunger, and seals
- New vent valve
- Pressurize the system >10,000 psi with flow results in stability and delta <20 psi using any combination of mobile phase
- Retention times are consistent and pressure holding stable during overnight runs; sensitivity good
ENVIRONMENT
- ACQUITY UPLC Binary Solvent Manager (BSM)
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
- ACQUITY UPLC I-Class BSM
CAUSE
Tubing and fittings leak.
Over time, service to pump head components during performance maintenance can contribute to a high leak rate and dynamic leak test failure.
FIX or WORKAROUND
- Replace the following tubing assemblies to manage suspected leaks at high-pressure fittings:
- Common pump head tubing and plumbing:
- 430001120 for the ASSY, TUBE, HEAD OUTLET TO XDUCER INLET
- 430001121 for the ASSY, TUBE, XDUCER OUTLET TO ACCUM CV INLET
- 430001773 for the ASSY, TUBE, XDUCER TO CHECK VALVE, i2V
- Pump A accumulator to vent valve
- 430001200 for the ASSY, TUBE, ACCUM "A" XDUCER - VENT VALVE (Port 3)
- Pump B accumulator to vent valve
- 430001199 for the ASSY, TUBE, ACCUM "B" XDUCER - VENT VALVE (Port 6)
- Common pump head tubing and plumbing:
ADDITIONAL INFORMATION
- If tubing is suspect, you may need to replace the following tubing assemblies:
- Pump A: 430001208 for the ASSY, TUBE, VENT VALVE PORT 2 TO TEE/FILTER
- Pump B: 430001207 for the ASSY, TUBE, VENT VALVE PORT 5 TO TEE/FILTER
- Pump outlet tubing: 430001486 for the TUBE ASSY, PUMP OUT, .010" FOR EVERFLOW INJ
- Loosely tightened/worn support plate can also cause excessive leak, part number: 700002601
