2695 Separations Module: What to do if negative peaks are detected irregularly - WKB192141
Article number: 192141
SYMPTOMS
- During a 20-minute run, negative peak appears irregularly, about twice
ENVIRONMENT
-
2695 Separations Module
- 2489 UV/Vis Detector
-
A-line: premixed, mobile phase is a water/acetonitrile/TFA mixture
-
Flow rate: 0.5 mL/min
-
Pressure: ~2700 psi; Δpsi: ~20
-
No negative peak observed with Stop Flow
CAUSE
Needle wash solvent contamination/air bubbles/TFA deterioration
FIX or WORKAROUND
- 0-µL injection, prep injection
→ Determine if the negative peak is caused by the injector motion.
0-µL injection: negative peak detected
Immediate injection: no negative peak detected
For the above result, there is a possibility of a malfunction around the injector and valve. Therefore, after changing the needle wash solvent to the same one as the mobile phase, perform prime and 0-µL injection (also, simultaneously acquire the system pressure data).
*This is a verification in order to confirm the possibility of needle wash solvent contamination. - Obtain system pressure
→ Check for possible affects from bubbles. - Prepare and run with a mobile phase without TFA.
→ When preparing the mobile phase, make sure to use TFA immediately after opening the ampoule.
To assess the possibility of degradation of TFA in mobile phase.
ADDITIONAL INFORMATION
