Skip to main content
Waters

2695 Separations Module: What to do if negative peaks are detected irregularly - WKB192141

Article number: 192141

SYMPTOMS

  • During a 20-minute run, negative peak appears irregularly, about twice

ENVIRONMENT

  • 2695 Separations Module

  • 2489 UV/Vis Detector
  • A-line: premixed, mobile phase is a water/acetonitrile/TFA mixture

  • Flow rate: 0.5 mL/min

  • Pressure: ~2700 psi; Δpsi: ~20

  • No negative peak observed with Stop Flow

CAUSE

Needle wash solvent contamination/air bubbles/TFA deterioration

FIX or WORKAROUND

  1. 0-µL injection, prep injection
    → Determine if the negative peak is caused by the injector motion.
    0-µL injection: negative peak detected
    Immediate injection: no negative peak detected
    For the above result, there is a possibility of a malfunction around the injector and valve. Therefore, after changing the needle wash solvent to the same one as the mobile phase, perform prime and 0-µL injection (also, simultaneously acquire the system pressure data).
    *This is a verification in order to confirm the possibility of needle wash solvent contamination.
  2. Obtain system pressure
    → Check for possible affects from bubbles.
  3. Prepare and run with a mobile phase without TFA.
    → When preparing the mobile phase, make sure to use TFA immediately after opening the ampoule.
    To assess the possibility of degradation of TFA in mobile phase.

ADDITIONAL INFORMATION

id192141, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent

Not able to find a solution? Click here to request help.