What is the plunger seal wash consumption on an Alliance separations module? - WKB19065

Article number: 19065

ENVIRONMENT

  • Alliance 2695
  • Alliance 2795

ANSWER

Each seal wash cycle 1 second, which translates to approximately 50 to 100 µL of wash volume consumed.

ADDITIONAL INFORMATION

Plunger Seal Wash period sets the time interval between successive plunger seal-wash pump cycles.

  • The plunger seal wash operates at a default value of 1 minute.
  • The value can be set to Off or 0.50 to 10.00 minutes in 0.01 minute increments.

 

2695_Alliance_seal wash.PNG

The seal wash period is part of the Alliance Default method set on the 2695 front panel—in mobile phase screen parameters > other parameters. 
You can set the interval.