Sample fluidics over pressure error on an I-Class SM-FTN when trying to prime syringes - WKB18810
Article number: 18810
SYMPTOMS
- When trying to prime the syringes, a sample fluidics over pressure error occurs
- The user can inject sequences without issues
- Able to prime when syringe inlet removed from inject valve
ENVIRONMENT
- ACQUITY I-Class UPLC system
- Sample Manager - Flow Through Needle (SM-FTN)
CAUSE
Clogged inject valve
FIX or WORKAROUND
- Remove the inject valve:
- Go to the ACQUITY/MS Console.
- Under the Sample Manager, Select Maintain, and then select Park Inject Valve.
- Remove the single screw at the 11-o'clock position; you can then pull out the inject valve.
- Place the inject valve face down in a beaker with enough solvent to cover the inside of the ports.
- Sonicate for 30 minutes and reinstall.
- If still blocked replace inject valve Part Number: 700006057
ADDITIONAL INFORMATION
id18810, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+