Skip to main content
Waters

Sample fluidics over pressure error on an I-Class SM-FTN when trying to prime syringes - WKB18810

Article number: 18810

SYMPTOMS

  • When trying to prime the syringes, a sample fluidics over pressure error occurs
  • The user can inject sequences without issues
  • Able to prime when syringe inlet removed from inject valve

ENVIRONMENT

  • ACQUITY I-Class UPLC system
  • Sample Manager - Flow Through Needle (SM-FTN)

CAUSE

Clogged inject valve

FIX or WORKAROUND

  1. Remove the inject valve:
    • Go to the ACQUITY/MS Console.
    • Under the Sample Manager, Select Maintain, and then select Park Inject Valve.
    • Remove the single screw at the 11-o'clock position; you can then pull out the inject valve.
  2. Place the inject valve face down in a beaker with enough solvent to cover the inside of the ports.
  3. Sonicate for 30 minutes and reinstall.
  4. If still blocked replace inject valve Part Number:  700006057

ADDITIONAL INFORMATION

id18810, A-30SM, UPISMFTN, UPPFTN, UPSMFTN, UPSMFTN+

Not able to find a solution? Click here to request help.