Alliance 2695 initialization failure - WKB18351
Article number: 18351
SYMPTOMS
- Alliance 2695 does not complete initialization
- Two beeps, sometimes followed by continuous beep
- LED screen is back-lit solid blue, gray, or "snowy"/pixilated but no text is visible
- Soft keys non responsive
- Instrument may attempt to initialize followed by a variety of initialization errors
- Instrument may intermittently power off
ENVIRONMENT
- 2695 Separations Module
- e2695 Separations Modules
CAUSE
CPU PCB failure
FIX or WORKAROUND
Contact Waters for a local service representative to replace this part.
Note: CPU replacement is considered a major repair.
ADDITIONAL INFORMATION