Sample set hangs up after the equilibrate line for ACQUITY UPC2 system - WKB17673
Article number: 17673
SYMPTOMS
- Sample set hangs up after the equilibrate line for ACQUITY UPC2 system
- The system equilibrates but does not continue to the subsequent injection
ENVIRONMENT
- ACQUITY UPC2 System
- Empower 3 Feature Release 4
- Driver Pack 2017
- Column Manager 30S
CAUSE
Defect
FIX or WORKAROUND
Add an equilibrate line to the sample set before each column switch. Set the run time to 10 seconds and the flow to 0.