What is the maximum pressure for the ACQUITY H-Class? - WKB17398
Article number: 17398
ENVIRONMENT
- ACQUITY H-Class
ANSWER
15,000 psi up to 1.0 mL/min, 9000 psi up to 2.0 mL/min (firmware version 1.5x and earlier)
15,000 psi up to 1.0 mL/min, 7800 psi up to 2.2 mL/min (firmware version 1.6x and later)