Error: Tray lost synchronization - WKB1523

Article number: 1523

SYMPTOMS

  • Needle pierces vial off center
  • Banging/grinding noise
  • Error state in console

ENVIRONMENT

  • H-Class
  • I-Class

CAUSE

Rotary tray out of calibration

FIX or WORKAROUND

  1. Calibrate the rotary tray, which is accessed through Service mode in the console.

ADDITIONAL INFORMATION

 

id1523, audible noise, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO