H-Class injection reproducibility is poor - WKB14800
Article number: 14800
SYMPTOMS
- H-Class injection reproducibility is poor and peak area is too small.
ENVIRONMENT
- ACQUITY UPLC H-Class SM-FTN
- ACQUITY UPLC I-Class SM-FTN
CAUSE
Six-way valve no.3, 4-channel is not free and suction is inaccurate.
FIX or WORKAROUND
The reproducibility is normal after replacing the six-way valve.
ADDITIONAL INFORMATION
Before replacing the six-way valve, confirm that the system leak test and sample syringe leak test passed.