Skip to main content
Waters

No power on the Alliance 2695 - WKB14166

Article number: 14166

SYMPTOMS

  • No power when powered on
  • Fans not running
  • No display on LCD display
  • Different power source and power cord does not work
  • Initialization fails - Screen goes blank then reboots

ENVIRONMENT

  • Alliance 2695

CAUSE

Bad Power Supply

FIX or WORKAROUND

Replace the power supply. See KB article 1739 - What is the part number for the power supply of an Alliance System?

ADDITIONAL INFORMATION

Not able to find a solution? Click here to request help.