Sample heater/cooler is not recognized on Alliance 2695 - WKB14051

Article number: 14051

SYMPTOMS

  • The sample heater/cooler is not showing as an installed option
  • A sparge valve shows as an installed option, but no such valve is present

ENVIRONMENT

  • 2690/2695/e2695
  • Alliance Separations Module

CAUSE

Bad connector port

FIX or WORKAROUND

Use one of the other three option connectors on the main CPU; J2, J3, J4, and J5 are all smart connectors.