How to connect ACQUITY UPLC Column Heater CM-A to ACQUITY UPLC Sample Manager SM-FTN - WKB13856

Article number: 13856

OBJECTIVE or GOAL

To connect the SM-FTN injection valve to the column select valve on the CM-A Column Manager.

ENVIRONMENT

  • ACQUITY UPLC H-Class Flow Through Needle (FTN)
  • ACQUITY UPLC I-Class Flow Through Needle (FTN)

PROCEDURE

  1. Install 700008869 - ASSY, WELDED TUBE, SS, .003 ID X 14.5 LG - to SM-FTN injector valve port 6.
  2. Install the other end of the tube to the central port of the Inlet Switching Valve on the CM-A Column Manager.