Area count %RSD fails company specification - WKB13325
Article number: 13325
SYMPTOMS
- Area count %RSD fails company specification; it usually passes easily
- Peak integration is fine
- PLUNO injection within range of 20% to 75% sample loop
- All lines primed
- Retention times OK
- Passes characterize needle seal and characterize needle and loop volume
ENVIRONMENT
- ACQUITY UPLC Sample Manager
CAUSE
Worn sample syringe
FIX or WORKAROUND
Replace the sample syringe (700002570 - Syringe 100 µL HP).
ADDITIONAL INFORMATION