ACQUITY Solvent Manager unable to compress during Leak Test - WKB12776

Article number: 12776

SYMPTOMS

  • Accumulator will not compress during leak test
  • Primary will not compress during leak test
  • no air in solvent lines
  • No visible leaks
  • Significant pressure fluctuations
  • Pressure fluctuation
  • Retention time shift
  • No flow from the pump

ENVIRONMENT

  • ACQUITY UPLC H-Class
  • ACQUITY UPLC
  • ACQUITY UPLC I-Class
  • nanoACQUITY UPLC
  • ACQUITY Arc
  • ACQUITY UPLC M-Class

CAUSE

Crack of plunger on primary and residue build-up on accumulator plunger.