With an ACQUITY PDA or a 2998 PDA, can I use the Events tab to switch the wavelength? - WKB126050

Article number: 126050

ENVIRONMENT

  • ACQUITY UPLC PDA Detector
  • ACQUITY UPLC PDA eλ Detector
  • 2998 Photodiode Array Detector
  • Empower 3

ANSWER

Yes, the option is available and can be done by first selecting the required single 2D channel on the 2D Channels tab and then setting the time to make the change in the Events tab, as shown below:

 

2998 2D Channels.JPG              2998 2D Events Channel.JPG

 

ADDITIONAL INFORMATION

Although this option exists, it is not required if using 3D collection; just extract the wavelength of interest.

Otherwise, you can collect data using a separate 2D channel. Empower allows for eight single 2D channels.