Skip to main content
Waters

Error message "Active Preheater not supported with current configuration" appears for ACQUITY UPLC H Class - WKB12412

Article number: 12412

SYMPTOMS

  • After switching between Column Heater - Active (CH-A) and 30cm Column Heater/Cooler module, this error message appears
  • Unable to start run with Column Manager - Active (CM-A) due to Columns "Active Preheater not supported with current hardware"

ENVIRONMENT

  • Empower
  • MassLynx
  • ACQUITY Column Heater - Active (CH-A)
  • ACQUITY Column Manager - Active (CM-A)
  • ACQUITY 30cm Column Heater - Active (CH-30A)
  • ACQUITY UPLC 30cm Column Heater/Cooler (30cm CHC)

CAUSE

  • A preheater conflict programmed in the ACQUITY Sample Manager instrument/inlet method and configured in console software
    • This issue can occur in a mixed column module environment when switching between 30cm column compartments (CH-30A and 30cm CHC)

FIX or WORKAROUND

Change the settings for the preheater (non-operating column compartment) in both the console and the instrument method.

  • The preheater will need to be configured back to active when switching operations back to this column module.

clipboard_e69596594331b6caf7174ccb7f8ae88f1.png

clipboard_eae268c508fbbeb794125c24aaef603c1.png

ADDITIONAL INFORMATION

 

Not able to find a solution? Click here to request help.