Error message "Active Preheater not supported with current configuration" appears for ACQUITY UPLC H Class - WKB12412
Article number: 12412
SYMPTOMS
- After switching between Column Heater - Active (CH-A) and 30cm Column Heater/Cooler module, this error message appears
- Unable to start run with Column Manager - Active (CM-A) due to Columns "Active Preheater not supported with current hardware"
ENVIRONMENT
- Empower
- MassLynx
- ACQUITY Column Heater - Active (CH-A)
- ACQUITY Column Manager - Active (CM-A)
- ACQUITY 30cm Column Heater - Active (CH-30A)
- ACQUITY UPLC 30cm Column Heater/Cooler (30cm CHC)
CAUSE
- A preheater conflict programmed in the ACQUITY Sample Manager instrument/inlet method and configured in console software
- This issue can occur in a mixed column module environment when switching between 30cm column compartments (CH-30A and 30cm CHC)
FIX or WORKAROUND
Change the settings for the preheater (non-operating column compartment) in both the console and the instrument method.
- The preheater will need to be configured back to active when switching operations back to this column module.