How to disassemble an ACQUITY UPLC FLR - WKB123889

Article number: 123889

OBJECTIVE or GOAL

Shut down and disassemble a FLR detector for shipment to the repair center.

ENVIRONMENT

  • ACQUITY UPLC H-Class FLR
  • ACQUITY UPLC I-Class FLR
  • ACQUITY UPLC FLR

PROCEDURE

  1. Wash the system with H2O for 20 minutes, and then wash the system with IPA for 20 minutes.
  2. Close the control software and power-off all modules on the stack.
  3. Electrically unplug the module and mark and remove any cables (Ethernet, signal, I/O, events) from the module.
    (Marking the cables is important for reinstalling the system when the modules come back from repair. Photos of cable and tubing are also helpful.)
    clipboard_eb21b987f5b851873f79e2629d56a7579.png
  4. Disconnect inlet and outlet tubing.
    clipboard_e6dfccacf8bb5406e491f6c2af38d3844.png
  5. Move the module to a suitable box for shipment. (If a box is not available, ask your local Waters office.)
    Note the dimensions and weight of the module to choose the right box:
    ACQUITY UPLC FLR 34.3 cm X 51.8 cm X 21.6 cm, 13.6 kg
  6. Fill and sign the latest revision of H&S module provided by Waters.
  7. Attach it to the box and send a PDF copy to the local Waters office.

  8. The Waters office will contact you to arrange the shipment.

ADDITIONAL INFORMATION

To reinstall the module, attach all of the cables and tubes that were removed.