Article number: 123799

SYMPTOMS

  • Peak retention times drift, appearing both early and late
  • Passes pump leak tests
  • Column temperature is ok and stable
  • System pressure is low

ENVIRONMENT

  • ACQUITY UPLC

CAUSE

Leak on the mixer

FIX or WORKAROUND

Replace the mixer. Confirm that system pressure increases and RT returns to normal.

ADDITIONAL INFORMATION