Carousel ID sensor fault error message for 2695 Separations Module - WKB12248

Article number: 12248

SYMPTOMS

  • Carousel ID sensor fault

ENVIRONMENT

  • 2695 Separations Module
  • Alliance HPLC System

CAUSE

The carousel's magnets are dirty or rusty, or the carousel carrier is defective.

FIX or WORKAROUND

  1. Change the carousel. (Part WAT270328 is a set of five carousels.)

ADDITIONAL INFORMATION

If enabled on the Configuration screen, the carousel ID sensor verifies carousel placement.

!HWERROR1! Check the magnet in the carousel.

If the error persists, replace the carousel/carrier drive (WAT270928).