The ACQUITY SM error "Missing/wrong plate or vial" is not triggered when vials are absent - WKB122043
Article number: 122043
SYMPTOMS
- Error "Missing/wrong plate or vial" is not triggered when vials are missing
ENVIRONMENT
- ACQUITY UPLC Sample Manager
- ACQUITY UPLC Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Using the "ANSI-48 Vials 2 mL Holder" in a sample set with plate "48 vials 2 mL holder" configured instead of "ANSI-48 Vials 2 mL Holder".
FIX or WORKAROUND
Set the ANSI-48 Vials 2 mL Holder on the sample set.
ADDITIONAL INFORMATION
48 vials 2 mL holder has a plate height of 25 mm, but ANSI-48 Vials 2 mL Holder has a plate height of 38 mm, so when the sample manager tries to inject the samples it finds the plate and the error is not triggered.
id122043, UPBINARY, UPIBSM, UPIBSM+, UPISMFL, UPISMFL+, UPISMFTIVD, UPISMFTN, UPISMFTN+, UPQSM, UPQSM+, UPQSMBIO, UPQSMBIO+, UPSMFTN, UPSMFTN+, UPSMFTNB+, UPSMFTNBIO