Unexplained injection delay with an Agilent 6890 GC and G1888 Headspace Sampler using Empower - WKB12177
Article number: 12177
SYMPTOMS
- Cables properly connected for GC with HSS
- Delay happens on two separate 6890/G1888 systems
- New instrument method created with the same delay
- No delay is set in the sample set
- GC oven time, GC cycle time, and run time are set properly
ENVIRONMENT
- Empower 3 Enterprise
- Agilent 6890 GC
- Agilent G1888 Headspace Sampler
CAUSE
Corrupted project
FIX or WORKAROUND
- Create a new project.
- Create a new instrument method.
ADDITIONAL INFORMATION
See article WKB62833 on split injection technique.
See article WKB62831 on splitless injection technique.
See article WKB62933 on PTV injection technique.
id12177, 6890A/+GC, 6890N GC, EMP2LIC, EMP2OPT, EMP2SW, EMP3GC, EMP3LIC, EMP3OPT, EMP3SW, EMPGC, EMPGPC, EMPLIC, EMPOWER2, EMPOWER3, EMPSW, HEADSPACE, SUP