Failed to obtain waters GI pump status on a nanoACQUITY - WKB12120
Article number: 12120
SYMPTOMS
- After software reload error "Failed to obtain waters GI pump state usage" is observed
ENVIRONMENT
- nanoACQUITY
CAUSE
nanoACQUITY System Driver not fully installed.
FIX or WORKAROUND
- Updating Software - Follow this procedure when updating software to 1.41:
- Delete the current system configuration in MassLynx.
- Power-off and restart the PC or select Stop ACQUITY Process.cmd.
- Power-off the instruments.
- Uninstall 1.30, 1.23b, or 1.22.
- Uninstall all microsoft.NET files, starting with the latest version.
- Go to Run and type "regedit".
- Delete the Waters folder under loc machine/software.
- Delete the Waters Instrument folder under Program files.
- Install nanoACQUITY 1.41.
- Power-off and restart the PC or select Stop ACQUITY Process.cmd.
- Power-on the instruments.
- Update firmware on all instruments.
- Open MassLynx and select the default project.
- Open the Inlet Method.
- Configure the system (Waters Pump Controller for Inlet and None for Autosampler).
- Power-off and restart the PC or select Stop ACQUITY Process.cmd.
- Power cycle the instruments (firmware load must be complete).
- Open MassLynx and open the Inlet Editor.
- Open the nanoACQUITY UPLC console and configure the system. (From console main menu, go to Configure/Configure instrument modules. Configure nBSM as the gradient BSM. Configure nBSM-2 as the optional BSM (first dimension or trapping BSM) when there are two nBSMs in a system. Click OK.)
- Close the console, Inlet Editor and MassLynx.
- Power-off and restart the PC. (if you skip this step, you will receive the following error message: "failed to obtain Waters GI Pump status".)
- Every time you change the system configuration in the console, you must power-off and restart the computer.
- Configure the system in MassLynx. (Select Waters Pump Control for Pump and None for Autosampler for all v1.41 system configurations.)
ADDITIONAL INFORMATION
id12120, UPNAYXPMP, UPNSAMMGR, UPNSOLMGR