Gradient does not start for Prep LC with manual injector - WKB121027
Article number: 121027
SYMPTOMS
- Gradient does not start in 2555 QGM when rotating manual injector from “Load” to “Inject”
ENVIRONMENT
- Prep 150 LC System
- Prep Manual Injector
- 2535/2545/2555 Quaternary Gradient Module
- 2489 UV/Vis Detector
- Waters Fraction Collector III
- Windows 10
- Empower 3 FR 4
- MassLynx
CAUSE
No, I/O cable to 2555Q.
FIX or WORKAROUND
Connect injector/start cable from manual injector to “Gradient Start” in 2555Q; “Inject start” in detector and WFCIII.
ADDITIONAL INFORMATION
To make manual injection in Empower:
1) Click Prepare.
2) When Empower indicates "Ready" and "Click Injection", turn the manual injector to the Inject position.
To make manual injection in MassLynx:
1) Start sample run.
2) When the inlet method indicates "Ready for injection", turn the manual injector to the Inject position. (Turning the injector too early results in no gradient triggering.)
