Injection valve outlet tubing leaks - WKB120949

Article number: 120949

SYMPTOMS

  • Low and unstable pressure
  • Solvent leaks at injection valve

ENVIRONMENT

  • ACQUITY UPLC Sample Manager
  • ACQUITY UPLC H-Class FTN
  • ACQUITY Arc

CAUSE

Damaged outlet tubing at the injection valve

FIX or WORKAROUND

  1. Remove the damaged outlet tubing.
  2. Install new outlet tubing.
  3. Verify that the injection valve stopped leaking at the outlet tubing.

ADDITIONAL INFORMATION