"Degasser H/W fault" with ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM) - WKB120923
Article number: 120923
SYMPTOMS
- "Degasser H/W Fault"
ENVIRONMENT
- ACQUITY UPLC H-Class Quaternary Solvent Manager (QSM)
CAUSE
The vacuum exhaust line is pinched in the drawer.
FIX or WORKAROUND
Remove the exhaust line from the drawer.
ADDITIONAL INFORMATION