No channel data available following method resolution utility on Agilent GC with ICF control - WKB119951
Article number: 119951
SYMPTOMS
- Instrument hardware configuration changed and method resolution triggered from Run Samples window > Edit Method
- Method changes saved and sample set run
- No channel data acquired
ENVIRONMENT
- Agilent 7890
- Agilent 8890
- ICF Control on Empower 3
CAUSE
The channel names are not updated by the method resolution utility.
FIX or WORKAROUND
- Return to Edit Method in the Run Samples window.
- Navigate to the Derived Channel Properties tab.
- Use the Channel ID drop-down menu to select the correct channel name.
ADDITIONAL INFORMATION
id119951, 7890 GC, 7890 GCDET, 7890BGC