Co-eluted peaks seen when specific batches of MassTrak AAA Eluent A and column are used together - WKB119759

Article number: 119759

SYMPTOMS

  • Co-eluted peaks seen when MassTrak AAA Eluent A Concentrate (part 186004090, lot 128100171) is used in conjunction with MassTrak Column of batch B0236
  • Tyrosine and lysine cannot be separated

ENVIRONMENT

  • ACQUITY UPLC BSM
  • MassTrak AAA

CAUSE

Root cause is unknown.

FIX or WORKAROUND

  1. Obtain a different lot of MassTrak AAA.
  2. Use a column of a different batch.

ADDITIONAL INFORMATION

Information was confirmed by GSS with PIM Chemistry & Consumables Team.

Both the items passed well within their individual specifications. However, when used together, the issue with co-eluted peaks is seen.