No solvent flow when wet priming 2695 Separations Module - WKB116127

Article number: 116127

SYMPTOMS

  • No solvent flow when wet priming 2695 Separations Module
  • Loosening outlet of gradient proportioning valve shows free flow of solvent
  • Loosening primary pump head tubing to check valve shows free flow of solvent

ENVIRONMENT

  • 2695 Separations Module
  • Empower

CAUSE

The check valve is stuck.

FIX or WORKAROUND

Replace the check valve (700000254).

ADDITIONAL INFORMATION

You can also try sonicating the check valves in 50/50 methanol/water for 30 mins as a temporary fix. 