No solvent flow when wet priming 2695 Separations Module - WKB116127

Article number: 116127

SYMPTOMS

  • No solvent flow when wet priming 2695 Separations Module
  • Loosening outlet of gradient proportioning valve shows free flow of solvent
  • Loosening primary pump head tubing to check valve shows free flow of solvent

ENVIRONMENT

  • 2695 Separations Module
  • Empower

CAUSE

The check valve is stuck.

FIX or WORKAROUND

Replace the check valve (700000254).

ADDITIONAL INFORMATION

You can also try sonicating the check valves in 50/50 methanol/water for 30 mins as a temporary fix. 

id116127, 2690, 2690D, 2690DUPGD, 2690UP, 2695, 2695D, 2695DE, 2695DEB, 2695DUP, 2695E, 2695EB, 2695INERT, 2695UP, A-10SSM, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, checkvalve, gradient proportioning valve