No chromatographic peaks - WKB115238
Article number: 115238
SYMPTOMS
- No chromatographic peaks in the expected time window
- No signal
- Peaks visible when acquiring from time 0
ENVIRONMENT
- ACQUITY UPLC
- Alliance HPLC System
CAUSE
Both solvent bottles (A and B) contain organic solvent.
FIX or WORKAROUND
Prepare the correct mobile phases.
ADDITIONAL INFORMATION
id115238, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent, UPBINARY