No chromatographic peaks - WKB115238
Article number: 115238
SYMPTOMS
- No chromatographic peaks in the expected time window
- No signal
- Peaks visible when acquiring from time 0
ENVIRONMENT
- ACQUITY UPLC
- Alliance HPLC System
CAUSE
Both solvent bottles (A and B) contain organic solvent or both solvent bottles (A and B) contain water or (A and B) lines swapped.
FIX or WORKAROUND
Ensure to use the correct mobile phases for the selected solvent lines.
ADDITIONAL INFORMATION
id115238, ALLCOLCLR, ALLCOLHTR, ALLCOLHTRB, eluent, UPBINARY