ACQUITY PDA Low Sensitivity - WKB111628

Article number: 111628

SYMPTOMS

  • Lower than expected sensitivity when injecting samples/standards

ENVIRONMENT

  • ACQUITY PDA

CAUSE

Dirty flow cell.

FIX or WORKAROUND

  1. Flush the flow cell with 25:25:25:25 - H20:MeOH:ACN:IPA

ADDITIONAL INFORMATION

 