ACQUITY PDA Low Sensitivity - WKB111628
Article number: 111628
SYMPTOMS
- Lower than expected sensitivity when injecting samples/standards
ENVIRONMENT
- ACQUITY PDA
CAUSE
Dirty flow cell.
FIX or WORKAROUND
- Flush the flow cell with 25:25:25:25 - H20:MeOH:ACN:IPA
ADDITIONAL INFORMATION