Faulty IEEE-488 cable leading to instrument failure in Empower - WKB11151
Article number: 11151
SYMPTOMS
- Getting a communications error in the Empower Message Center
- "System Idle Instrument Failure" observed
- There is no conflict between IEEE address selection and the other instrument modules in the system
- Instrument method and system configuration match up properly
- IEEE cables are connecting properly. There is no dust, moisture, or corrosion in the connectors.
ENVIRONMENT
- Empower Software
- All instrument hardware using IEEE-488 communications
CAUSE
- Faulty IEEE Cable or Connection
- Reseat Connections
- Swapping known good cable directly to the failing instrument module isolated the problem to a bad cable
FIX or WORKAROUND
- Reseat Connection
- Replace the IEEE Cable