Liquid collects under the sample manager - WKB110590
Article number: 110590
SYMPTOMS
- Liquid puddles under the sample manager
- No visible leaks
- Running the defrost procedures fails to resolve issue
- Shimming the back feet of the sample manager fails to resolve issue
ENVIRONMENT
- ACQUITY UPLC Sample Manager - Fixed Loop (SM-FL)
- ACQUITY UPLC H-Class Sample Manager - Flow Through Needle (SM-FTN)
- ACQUITY UPLC I-Class SM-FTN
- ACQUITY UPLC I-Class SM-FL
- ACQUITY UPLC M-Class SM-FL
CAUSE
Blockage in drain line
FIX or WORKAROUND
Replace the drain tubing or clear the blockage.
ADDITIONAL INFORMATION
This image indicates the cause and most probable leak location on the ACQUITY Classic UPLC Sample Manager:

