Poor injection reproducibility with ACQUITY UPLC H-Class PLUS Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS) - WKB109495
Article number: 109495
SYMPTOMS
- Poor reproducibility from injection to injection
- Random drops in sample manager pressure during injection cycle
- "Sample Draw Rate Exceeded" messages
- Correlation observed between sample pressure drops in Sample Manager - FTN+ and low peak response in chromatogram
ENVIRONMENT
- ACQUITY UPLC H-Class PLUS Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS)
CAUSE
Air in the sample manager fluidics lines
FIX or WORKAROUND
Prime the ACQUITY UPLC H-Class PLUS Sample Manager - Flow Through Needle PLUS (SM-FTN PLUS) from the ACQUITY Console.
ADDITIONAL INFORMATION